Multifunction Process Calibrator  YHS-724

  • Easy to read measure/source screen lets you view input and output simultaneously
  • Measure volts, mA, RTDs, thermocouples and ohms to test sensors and transmitters
  • Source/simulate volts, mA, thermocouples , RTDs, frequency, ohms, and pressure to calibrate transmitters
  • Support flow meter testing with frequency and CPM functions
  • Perform fast linearity tests with auto- step and auto-ramp features
  • Power transmitters during test using loop supply with simultaneous mA measurement
  • Handling of fast pulsed RTD transmitters and PLCs, with pulses as short as 10 ms
  • Store frequently used test setups for later use
  • Large battery capacity of four AA cells ;Battery door for easy changes
  • A split-screen display. The upper display allows users to measure volts, current, and only. The lower display allows the user to measure and source volts, current , resistance temperature detectors, thermocouples and ohms.
  • A thermocouple (TC) input/output terminal with cold junction temperature compensation, ambient temperature can be set and calibrate within the menu function.
  • Each parameters can manual and automatic stepping and ramping output, signal output can be arbitrarily set within the 0% and 100% of the value.
  • In the configuration menu can be set manually : Power off time for Instrument, backlight, off time, Screen Contrast, Cold junction compensation ON/OFF, HART, resistance, ON/OFF and ETC.
  • Have calendar clock function, it can be calibrated in configuration menu.
  • It can be store 8 sets of measurement and output data and call the showing at any time, all the settings and the functions of the last operation be auto-stored after power off.
  • High capacity rechargeable battery pack, with power adapter, within integrated charge-management functions.
  • Novel shape, large screen dot-matrix graphic display
  • Loop Power Voltage: 24V
  • Accuracy: 10%
  • Maximum Current: 24 mA, Short-circuit protection

Measure Accuracy:

Function

Measure Singal

Measure  Accuracy

 

DC Voltage

0.000—30.000 V

0.02%+3 counts  (The upper display)

0.000—24.000 V

0.02%+ 3 counts (The lower display)

0.000—90.000 mV

0.01%+ 10 counts

-10.000 —80.000mV 0.01 % + 10 counts (use the thermocouple jack )

DC Current

0.000—24.000 mA

0.02%+ 4 counts(The upper display)

0.000—24.000 mA

0.02%+ 4 counts(The lower display)

Resistance

0.00— 400.00Ω

0.1Ω (4 wire), 0.15Ω (2 wire and 3 wire)

400.0—4000.0Ω

0.5Ω (4 wire), 1Ω (2 wire and 3 wire)

Output Accuracy:

Output

Range

      Accuracy

DC Voltage

0.000— 20.000 V

0.02%+3 counts

0.000—90.000 m V

0.01%+10 counts

DC Current

0.000—24.000 mA

0.02%+ 4 counts

Resistance

5.00 — 400.00 Ω

0.15 Ω (Excitation current 0.15 to 0.5 mA), 0.1Ω (Excitation current 0.5 to 2 mA)

400.0 —4000.0 Ω

0.5Ω(Excitation current 0.05 to 0.8 mA)

RTD and Thermocouple:

Type

Measure Accuracy

Output Accuracy

PT-100 (385)

0.3 °C(4 wire ), 0.5 °C(2 wire and 3 wire)

0.3 °C

PT-100 (3926)

0.3 °C(4 wire ), 0.5 °C(2 wire and 3 wire)

0.3 °C

PT-100 (JIS)

0.3 °C(4 wire ), 0.5 °C(2 wire and 3 wire)

0.3 °C

PT-200 (385)

0.8 °C(4 wire ), 1.6 °C(2 wire and 3 wire)

0.8 °C

PT-500 (385)

0.4 °C(4 wire ), 0.9 °C(2 wire and 3 wire)

0.4 °C

PT-1000 (385)

0.2 °C(4 wire ), 0.3 °C(2 wire and 3 wire)

0.3 °C

Cu50

0.4 °C(4 wire ), 0.6 °C(2 wire and 3 wire)

0.4 °C

Cu100

0.4 °C(4 wire ), 0.6 °C(2 wire and 3 wire)

0.4 °C

J:

0.5 °C

0.7 °C

K:

0.8 °C

0.8 °C

T:

   0.5 °C

0.5 °C

E:

0.5 °C

0.5 °C

R:

1.5 °C

1.0 °C

S:

1.5 °C

1.5 °C

B:

1.8 °C

1.0 °C

N:

0.8°C

0.8 °C